CLA-FL2( High Performance Chemiluminescense Analyzer, Flow type)

 

 



       
1. Features                                            

        고감도 측정 (10-14 to 10-15 Watt/cm2S)
        HPLC용 Flow cell만 장착 가능
        플로우 인젝션 시스템 (FIA) 및 HPLC system 연동가능.
        Pen recorder, Integrator 또는 Chromatocoder에 대한 아날로그 출력 장치제공.
        샘플챔버의 온도를 50C까지 가열가능

       
 


 

 

 

 


2. Applications

  분석화학
   과산화수소, 폴리페놀 등 미량물질의 고감도 분석

 

 

 

 

 

 

 

 


 

 



3.Specifications                                               

 

 

 

Counting

  Single Photon counting (SPC)

Detection Wavelength

  300nm - 650nm(max.quantum efficiency: 430nm)

Sensitivity

  Approximately 10-14 to 10 -15 watt/cm2S (at 400nm to 500nm)

 Cooling

  PMT is Cooled by Peltier elements
  The Peltier elements are in turn cooled by circulating water

 Temp. Controller

  Operated manually or automatically.

 Heating Range

  Ambient Temperature to 50C

Counting Value

  Indicated on the LED
  (the counting stops as it reaches 1,000,000cps. For protecting   the detection device against damage from strong light.)

Output

  Recorder Output:1.0V/F.S / Integrator Output:1.0V/F.S

Function

  Gain/Attenuation/Auto Zero/Marker/Alarms

 AC Source

  AC100V,50/60Hz,5A

Size/Weight

  310(W)x420(D)x524(H)mm / approximately 35kg

 

 

 

 

 

 

 

 

 

 

제목 없음